1.4.8 Calibration point 2
Choose the measured value in physical unities - beginning and end of
measurement range.
1.4.9 Measured value 2
Input value is collected and assigned to the shown measured value.
1.4.10 Calibration point N
. . . depends on the number of calibration points
1.4.11 Measured value
Input value is collected and assigned to the shown measured value.
For 2.1 to 2.4 / 3.1 to 3.4 / 4.1 to 4.4: analogue to point 1
5. Analogue output
With and you make a choice
according to the display. With you quit
the menu point without any change and with
the input is accepted and you arrive
1 menu point lower to make your settings.
5.1 Beginning of range
Choose the output in the range of 0 ... 22 mA
(Standard 4 mA).
With you set the value to 0.0. With the
numbers you input the beginning value and
with the input is accepted and you quit the
menu level.
5.2 End of range
Choose the output in the range of 0 ... 22 mA
(Standard 20 mA).
With you set the value to 0.0. With the
numbers you input the beginning value and
with the input is accepted and you quit the
menu level.
5.3 MIN limit
Choose the minimal output value in the range
of 0 ... 22 mA (Standard 3 mA).
With you set the value to 0.0. With the
numbers you input the beginning value and
with the input is accepted and you quit the
menu level.
C
8
£
¤
C
8
C
8
C
8
5. Analogue output
£
5.1 Beginning
4.0
mA
¤
5.2 End 20.0 mA
5.3 Minimum 0.3 mA
C
5.4 Maximum 21.0 mA
8
q
Analogue output
7 8 9
Range beginning
4 5 6
4.0
mA
1 2 3
C
0
8
Analogue output
7 8 9
Range end
4 5 6
20.0
mA
1 2 3
C
0
8
Analogue output
7 8 9
MIN limit
4 5 6
3.0
mA
1 2 3
C
0
8