13
6. Commissioning
At the first commissioning of the M-Sens 2 it is necessary to calibrate the sensor.
Please consider:
•
Correct connection between sensor, C1-box and transmitter.
•
Correct installation of the sensor in respect to the wall thickness.
In case of negative results in spite of consideration of the points as stated above, please contact SWR.
Commissioning of the M-Sens 2
After the delivery the sensor is
not calibrated to the product(s),
so each calibration and parameterization
has to be executed during the commissioning. Therefore it is necessary to assign the measured moisture to
the desired display and to the initial value. The menu functions are very self-explanatory. In the following a
short introduction to the menu overview:
Changed values are confirmed and saved by pushing the arrow button. To abandon the menu without saving
the changes push the C-button.
Starting the menu
Entering the menu takes place by pushing 5 s the invisible key lower
right corner of the touch screen panel.
Basic function
For the measurement and display of the moisture in %, there 2 - 3 moisture
measurement executed with known and different moisture values. After having
entered the according moisture values by the panel, the system calculates the
moisture in % on the basis of those measurement values.
Hereby the basic function is initiated.
Adjustment
The menu point 1 “products” can be used for every of the 4 possible products.
Alarm
Alarm can be determined for every product in the menu point “measurement
range”.
Analogue outputs
The analogue outputs can be defined in menu point 5. They can be adapted to
the operator’s requirements (e. g. 0 - 20 mA).
The output value (current) is assigned to the measurement range.
Standard MIN = 4 mA
MAX = 20 mA
The measurement range filter serves to the adjustment to slower measuring
devices or to continuous output at the analogue output.