AMI Codes-II
Default Values
A-96.250.571 / 050517
97
11.
Default Values
NOTICE:
The parameter Cleaning is only visible if an optional
Cleaning Module is connected to the AMI Codes II.
The parameters pH and Temperature are only visible if the
option pH is installed.
Operation:
Sensors:
Filter Time Const.: .................................................................30 Sec
Hold after Cal.:.....................................................................120 Sec
Meas. Interval: ......................................................................... 5 Min
Default pH
(If pH option is not installed)
.............................. 7.00 pH
Alarm Relay
...................................................................... same as in Installation
Relay 1and 2
...................................................................... same as in Installation
Input
...................................................................... same as in Installation
Logger:
Logger Interval:........................................................... Event Driven
Clear Logger:................................................................................ no
Installation:
Sensor:
Disinf:........................................................................... Free chlorine
Dimension.................................................................................. ppm
Interpolation:............................................................................... yes
Ref. Verification: ...................................................................... 0.255
Standard: Standard 1: ......................................................... 7.00 pH
Standard: Standard 2: ......................................................... 9.00 pH
Cleaning ......................................................................... 2 Solutions
Signal Output
1&2
Parameter:................................................................................. DIS.
Current loop:...................................................................... 4–20 mA
Function:.................................................................................. linear
Scaling: Range low:........................................................... 0.00 ppm
Scaling: Range high: ......................................................... 5.00 ppm
Scaling: pH: Range low: ..................................................... 0.00 pH
Scaling: pH: Range high:................................................... 14.00 pH
Scaling: Temperature: Range low:......................................... 0.0 °C
Scaling: Temperature: Range high: ...................................... 50.0 °C
Scaling: Sample Flow: Range low: ...........................................0 B/s
Scaling: Sample Flow: Range high:......................................200 B/s
Alarm Relay
Alarm DIS.:
Alarm high: ........................................................................ 5.00 ppm
Alarm low:.......................................................................... 0.00 ppm
Hysteresis:......................................................................... 0.10 ppm
Delay: ......................................................................................5 Sec