![Swann AMI Codes-II Скачать руководство пользователя страница 50](http://html1.mh-extra.com/html/swann/ami-codes-ii/ami-codes-ii_operators-manual_1411948050.webp)
48
A-96.250.571 / 050517
AMI Codes-II
Maintenance
6.3.2
Reagents for Measuring Monochloramine and Ozone
2 L of reagents last for 15 days with 2 min. interval:
1 x 50 ml of Oxycon On-line DPD
1 x 240 gr of Oxycon On-line Buffer
1 x 60 gr of Reagent Oxycon On-line KI
4 liters of demineralized water
Prepare
Oxycon
On-line DPD
1
Rinse the canister [E] labelled OXYCON ON LINE DPD-Re-
agent with demineralized water.
2
Fill the canister up to the 2 liter mark with demineralized water.
3
Slowly pour the content of a bottle of concentrate Oxycon On-
line DPD into the canister.
Avoid splashing!
4
Close the canister with the screw cover and tighten it well.
5
Mix the demineralized water and the reagents well..
6
Put the canister [E] into the holder [G].
7
Remove the screw cover and insert the suction lance [A] and
tighten the screw cover [B].
Prepare
Oxycon
On-line Buffer
& KI
1
Rinse the canister [F] labelled “OXYCON ON LINE Buffer” solu-
tion with demineralized water.
2
Fill the canister up to the 2 liter mark with demineralized water.
3
Slowly pour the content of one bag of Oxycon On-line Buffer
into the canister.
Avoid splashing!
4
Add the content of one bag Oxycon On-line KI to the same can-
ister.
5
Close the canister with the screw cover and tighten it well.
6
Mix the demineralized water and the reagents well.
7
Put the canister [F] into the holder [G].
8
Remove the screw cover and insert the suction lance [A] and
tighten the screw cover [B].
9