![Swann AMI Codes-II Скачать руководство пользователя страница 81](http://html1.mh-extra.com/html/swann/ami-codes-ii/ami-codes-ii_operators-manual_1411948081.webp)
AMI Codes-II
Program List and Explanations
A-96.250.571 / 050517
79
3.4 Set Time
Adjust date and time.
3.5 Cleaning
Automatic cleaning process using the Cleaning module-II
controlled by transmitter of AMI Codes-II.
Measurement is interrupted.
Flow error interrupts cleaning.
Cleaning does not start: E022 Reagent empty, E023 Cleaning
solution, no flow.
3.5.1
Parameter:
3.5.1.1
Mode: Select daily, weekly or off.
3.5.1.xx
Start time /
Calendar
: Time off day respectively weekday(s) for the
automatic start of the cleaning process.
3.5.1.3
Delay: During cleaning plus the delay time, the status of the signal-
and control outputs is as get below.
Range: 60–6’000 sec
3.5.1.4
Signal Outputs: Select the operation mode of the signal outputs
during cleaning:
3.5.1.5
Output/Control: Relay or signal output:
3.5.2
Fill Channel 11: Activates the cleaning pump and switches the
valve to the cleaning solution 1 (right canister)
3.5.3
Fill Channel 12: Activates the cleaning pump and switches the
valve to the cleaning solution 2 (left canister)
Cont.:
Signal outputs continue to issue the measured value.
Hold:
Signal outputs hold the last valid measured value.
Measurement is interrupted. Errors, except fatal errors,
are not issued.
Off:
Signal outputs are switched off (set to 0 or 4 mA).
Errors, except fatal errors, are not issued.
Cont.:
Controller continues normally.
Hold:
Controller continues based on the last valid value.
Off:
Controller is switched off.