![Swann AMI Codes-II Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/swann/ami-codes-ii/ami-codes-ii_operators-manual_1411948077.webp)
AMI Codes-II
Program List and Explanations
A-96.250.571 / 050517
75
9.
Program List and Explanations
1 Messages
1.1 Pending Errors
1.1.5
Provides the list of active errors with their status (active,
acknowledged). If an active error is acknowledged, the alarm relay
is active again. Cleared errors are moved to the Message list.
1.2 Maintenance List
1.2.5
Demands necessary maintenance, e.g. preparing new reagents.
1.3 Message List
1.3.1
Shows the error history: Error code, date and time of issue and
status (active, acknowledged, cleared). 65 errors are memorized.
Then the oldest error is cleared to save the newest error (circular
buffer).
2 Diagnostics
In diagnostics mode, the values can only be viewed, not modified.
2.1 Identification
Designation
:
View the Designation of instrument.
Version
: Firmware of instrument (e.g. V6.20-08/16)
2.1.3
Peripherals
: PeriClip 1: Firmware of peristaltic pump (e.g. 1.03)
2.1.4
Factory Test
: Test date of the Instrument, Motherboard and
Frontend. QC factory test.
2.1.5
Operating Time
: Years/Days/Hours/Minutes/Seconds
2.2 Sensors
2.2.1
Photometer:
Current value
: Shows the actual photometer signal [ppm].
Raw value
: Shows the actual photometer signal [Hz].
Absorbance
: Process value, depends on sample.
2.2.1.4
Cal. History:
Shows the diagnostic values of the last calibrations.
Number
: Calibration counter.
Date, Time
: Date and time of the calibration.
Slope
: Slope is a correction factor calculated on the basis of a
process calibration.
Range: 0.8–1.2