![Swann AMI Codes-II Скачать руководство пользователя страница 63](http://html1.mh-extra.com/html/swann/ami-codes-ii/ami-codes-ii_operators-manual_1411948063.webp)
AMI Codes-II
Maintenance
A-96.250.571 / 050517
61
6.10.2 Replace the Reagent Tubes
Tube
numbering
Level
1
2
1
2
3
J
K
Q
4
Nr.
from
to
1
Pump outlet: rear frame
Flow cell block: connection 1
see Flow cell block side view Q
2
Pump outlet: front frame
Flow cell block: connection 2
see Flow cell block side view Q
3
Reagent canister (J)
Oxycon on-line DPD
Pump inlet: rear frame
4
Reagent canister (K)
Oxycon on-line Buffer / KI
Pump inlet: front frame