Maintenance
6.3.
Refill or replace Reagents
The liquid level in canister 2 is monitored. The following messages
are displayed:
NOTICE:
Before refilling the reagents, rinse the canisters with
demineralized water.
Reagent
consumption
The 2 liter reagent canister will last for 15 days of operation (with
default measurement interval of 2 minutes). The provided reagent
set (for 8 canisters) therefore lasts for 4 months of operation.
As the reagent consumption is not linear find further examples be-
low:
WARNING
Health hazard
Some reagents can cause severe burns or eye damage.
For safe handling of the reagents you must read and under-
stand the Material Safety Data Sheets (MSDS), see
.
Canister almost
empty
Maintenance E065 - Reagents low and the
remaining reagent volume in %
(starting at 17 % = 340 ml).
Canister empty
Error E022 - Reagent empty
Measuring interval
Duration per
canister
Duration per
reagent set
1 minute
~ 7 days
2 months
2 minutes
~ 15 days
4 months
4 minutes
~ 22 days
5 months
6 minutes
~ 33 days
8 months