48
SVANTEK 974 User Manual
6.2 Setting up the units and scale of results presentation - Display Scale
The
Display Scale
sub-list enables the
user to define the result units (absolute or
logarithmic), adjust scale of plots and
switch the grid on/off.
<ENT>
Scale of results presentation
The
Scale
position defines the units of
results:
Lin
(linear) and
Log
(logarithmic).
In case of
Log
the graphical presentation is
given in the logarithmic scale and the
measurement results are expressed in
decibels (the result is related to the values
set in the
Reference Levels
window (
path:
<Menu>
/
Auxiliary
Setup
/
Reference Levels
).
=>
=>
Note: The parameters can be presented in Logarithmic (dB) or Linear (for example, m/s
2
) units. It
depends on the Scale parameter value. For example, 10 m/s
2
can be presented as 140 dB.
The
Resolution
position defines the
number of digits after the decimal point in
the presented results: one digit after the
decimal point (
0.1 dB
) or two digits after the
decimal point (
0.01 dB
).
►
Scaling the vertical axis of the plot
The
Dynamic
position enables the user to
select the required dynamic range scaling
of the plot. It is possible to select the range
from the set:
10dB
,
20dB
,
40dB
,
80dB
,
100dB
and
120dB
.
=>
=>