SVANTEK 974 User Manual
12.5. Firmware upgrade
SVANTEK is committed to continuous innovation path of development, and as such reserves the right to provide
firmware enhancements based on user’s feedback.
To update the instrument firmware:
•
Unpack the provided firmware package. (provided as a suitable compressed file).
•
Make sure the unit is turned off.
•
Connect SC 16 cable to the computer and SV 974 instrument (USB interface).
•
Keeping pressed the
<ENTER>
and
▲
push-buttons switch on the instrument - the following message
should appear on the unit's screen: BOOTSTRAP v1.06 (or higher).
•
Wait for the message
<USB>
on the unit’s screen and start from the PC:
go-usb.bat
.
•
The changing number and final message: "..... o.k." should appear on the computer screen.
•
Successful firmware update will be indicated by the message:
Program loaded!
•
Switch off the instrument.
Note: With the use of Supervisor or + software it is very easy to check if there are any
new firmware releases available for download.
12.6.
Storing the instrument
•
To preserve the life of the internal batteries, it is recommended that the instrument is turned off when
it is stored. In case of alkaline batteries, it is recommended to extract them out of the instrument.
•
Do not store the instrument permanently connected to the USB port. It shortens battery lifecycle.
•
When the instrument is turned off, it still draws a small amount of battery power. Therefore,
it is recommended to charge the cell every few months if it is not going to be used regularly.
12.7.
Transportation and carrying
For transportation or storage purpose, we recommend using the packaging provided by the manufacturer. In a
potentially dirty industrial environment it is advisable to use the carrying case provided by the manufacturer such
as the fabric material case (
SA 47
), lightweight case (
SA 143
)
or waterproof case
(
SA 79
), which ensures
excellent mechanical and environmental protection and long term storage conditions.
12.8.
Cleaning
Clean the surface of the instrument with damp soft cloth.
The instrument sockets should be cleaned with the use of compressed air.
Note: In cases of larger dirt, such as oil or grease, contact your Local Authorized Distributor or
Svantek Service Office.
12.9.
Troubleshooting
•
In case your instrument does not respond proceed with hardware reset of the instrument (see chapter
12.4).
•
In case the reset does not help call your Local Authorized Distributor or Svantek Service Office.
Should your SVANTEK professional measurement equipment need to be returned for repair or for calibration,
please contact the service office at the following number or contact via the SVANTEK’s website.