The
Logging
list enables the user to program
the logger functionality - recording of the
measurement time history results; and
program markers and parameters of the
signal recording (event or wave). The
Logging
list consists of seven positions:
Logger Setup
,
Logger Results
,
Summary Results
,
Logger Trigger
,
Event Recording
,
Marker Setup
and
Wave Recording
.
<ENT>
5.4.1 Setting up the logger general parameters
– Logger Setup
The
Logger Setup
list enables the user to
activate logger functionality. If
Logger
position
is
switched
off
only
Summary Result
s and
Wave Recording
positions are accessible in the
Logging
list.
<ENT>
Note: If Logger is Off, logger files are not created automaticallyand measurement results of the
time history changes are not saved!
The
Logger
position switches on the
functionality, which enables the user to save
selected results from the three profiles and
spectra with the defined interval selected in the
Logger Step
position.
►
The
Summary Results
position allows the user to select or deselect the saving
of the full set of profile results that the instrument generates during total
measurement time and which are not belonged to the time history data. These
results are:
RMS
,
Ovl, Peak
,
P-P
,
MTVV
as well as spectra in the analyzer
function. In addition to above mentioned results it is possible to add additional
data from the
Summary Results
list.
The
Logger Step
defines the step of the data logging in a file. It can be set from
2 milliseconds to 1 second or the values from 1 second to 59 seconds or the
values from 1 minute to 59 minute and up to 1 hour.
The
Logger Name
position enables the user
to define the logger file name. The default
name is
L
. The name can be up to eight
characters long. After pressing the
◄ or ►
push-buttons, the special window with text
editing function is opened.
►