SVANTEK 974 User Manual
6. SETTING UP THE DATA VIEW
– Display
The
Display
list contains the elements that
enable
the
user
to
independently
programme the display parameters.
The content of the
Display
list is different
for the spectrum analyser functions
(
1|1 Octave
,
1/3 Octave
and
FFT
) in
regards to the
Level Meter
function.
<ENT>
The
Display
list is used for setting the various parameters, which are dedicated to the control of the LCD screen
display, and contains the following items:
Display Modes
enables the user to select the mode of the measurement results presentation;
Display Scale
enables the user to change the scale in the graphical modes of result’s presentation;
Logger View
enables the user to select and present the results stored in the logger’s files;
Spectrum View
enables the user to change the type of the spectrum and to activate the
Max
and
Min
spectrum. This position appears only in spectrum analyser functions;
Spectrum Type
enables the user to change the spectrum type presented on the display:
Acceleration
,
Velocity
and
Displacement
. This position appears only in spectrum analyser functions;
Screen Setup
enables the user to set up the brightness and the switch on/off the energy saver function.
6.1 Selection of the view modes - Display Modes
The One Result view is always available in
all measurement modes. Other view modes
can be switched on or off in the
Display Modes
sub-list.
The mode of the results presentation is
related to the selection of the instrument’s
function (
VLM
,
1/1 Octave
,
1/3 Octave
,
etc.).
<ENT>
In case of
Level Meter
, following optional views are available:
1 Prof. & 3 Prof.
,
1 Prof. & Logger
,
3 Prof. & LoggerB,
Logger
and
File Info
.
One Result view
The One Result view is always available in all measurement modes. In this view, only one measured value is
displayed in large letters.
Field description of the One Result view
1.
File name when
Logger
is active (
path: <Menu> /
Measurement / Logging / Logger Setup /
Logger:
)
2.
Profile number
3.
The value of the measured function
4.
The name of the implemented filter:
HP
,
HP1
,
HP3
,
HP10
,
Vel1
,
Vel3
,
Vel10
,
VelMF
,
Dil1
,
Dil3
,
Dil10
,
Wh
5.
Detector type:
Lin
when the
RMS Integration
is
linear
(path:
<Menu>
/
Measurement
/
General Settings)
or:
100 ms
,
125 ms
,
.. 10.0 s
,
6
1
3
7
5
4
8
2