SVANTEK 974 User Manual
The
Filter
position displays the used
weighting filter (
HP
) during event signal
recording.
The
Trigger on Marker
position switches on or off the triggering by marker.
When
Trigger on Marker
is switched on then event recording will start by
initiation of one of the user controlled markers. Markers for triggering are defined
in the
Markers Setup
window.
In the
Trigger
position the following options
are available:
Off
,
Slope +
,
Slope
–
,
Level +
,
Level
–
,
External
and
Integr. Period
.
All triggers except
Integr. Period
were
described in the chapter 5.2.
►
…
Note:
When event recording is waiting for the slope trigger the “trigger
slope” icon appears alternatively with the „note” icon.
or
Note:
When event recording is waiting for the level trigger the “trigger
level” icon appears alternatively with the „note” icon.
or
Note: When event recording is waiting for the
“external” or
“integration period” trigger the “trigger” icon appears alternatively with
the „note” icon.
If the
Integr. Period
trigger was defined, the signal is recorded every time the
measurement starts and ended after the
Integration Period
time, defined in the
General Settings
window (
path: <Menu> / Measurement / General Settings
).
The
Source
position indicates the triggering signal source. Only one measured result can be used as a source
of the triggering signal in all modes, namely the output signal from the RMS detector coming from the first profile
which is denoted here as
RMS(1)
. This position does not become active.
The level of triggering signal for the event recording
(
Level
) can be set with 1 dB steps in the range from
1mm/s
2
(60 dB) to 10 km/s
2
(200 dB). The level can
be expressed not only in linear units, but also in
decibels. The vibration unit scale can be set in the
Display Scale
window (
path: Menu / Display /
Display Scale / Scale
).
In the
Trigger Period
position, it is possible to select the time interval of
checking the triggering conditions. This parameter can be set at:
Log. Step
(same as
Logger Step
value),
0.5ms
,
100.0ms
and
1s
.
In the
Rec. Time
position, it is possible to select the time of signal recording after
triggering. If another triggering condition that satisfies the selected conditions
appears, then the signal will be recorded for an additional
Rec. Time
. The
available values can be selected from
1s
to
8h
.