38
SVAN 977C User Manual
4.
Switch on the calibrator (if the used calibrator doesn’t have auto run function)
and wait approximately 30 seconds for the tone to stabilise before starting
the calibration measurement.
5. Start the calibration measurement by pressing the
<Start/Stop>
key.
The instrument performs series of 1-second consecutive measurements
displaying levels of the measured calibration signal.
If maximal difference between three 1-second consecutive results (Leq(C)) is
less than
0.05dB
, the calibration measurement will be stopped, and the
calibration factor calculated. The measurement can be always stopped by the
<Start/Stop>
key.
After calibration measurement stop,
Calibration Result
(measured reference
signal without calibration factor correction) and
New Calibration Factor
(difference between the Calibration Level and the Calibration Result, calculated
in dB) are displayed.
6. Press
<ENTER>
to save the new calibration factor (
Enter to Confirm
), or press
<ESC>
to reject it. In both
cases the instrument exits the
Calibration by Measurement
screen.
It is recommended to repeat the calibration measurement few times. Obtained results should be almost the same
(with
0.1 dB difference). Reasons for unstable results are as follows:
•
the calibrator is not properly attached to the instrument,
•
there are external acoustic disturbances such as high noise levels close by,
•
the calibrator or the measurement channel (the microphone, the preamplifier or the instrument itself) are
damaged.
7. Detach the calibrator from the microphone.
Note: During the calibration measurement, any external disturbances (acoustic noise or vibrations)
should not exceed a value of 100 dB (when using a calibrator that generates a level of 114 dB).
Note: Maximum deviation of the calibration factor cannot exceed +/-1.5 dB. In other case please
contact Svantek service.
3.3.5 Calibration by Measurement (Vibration signal)
Calibration by measurement of the vibration signal can be done in the following
way:
1.
Open the
Calibr. by Measurement
position in the
Calibration
list.
2.
Set the calibration level (
Cal. Level
) appropriate to the used calibrator. The
default level for calibration in the vibration mode is 10 m/s
2
at 159.2 Hz.
Remember to change this level if using an alternative vibration calibration
signal source.
3.
Attach the instrument’s accelerometer to the vibration calibrator using an
appropriate or recommended fixing method.
4.
Switch on the calibrator and wait approximately 30 seconds before starting
the calibration measurement.
5.
Start the calibration measurement by pressing the
<Start/Stop>
key.
Содержание Shawcity SVAN 977C
Страница 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Страница 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Страница 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Страница 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 250: ...250 SVAN 977C User Manual Appendixes...
Страница 262: ...262 SVAN 977C User Manual Appendixes 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter...
Страница 263: ...263 SVAN 977C User Manual Appendixes 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter...
Страница 264: ...264 SVAN 977C User Manual Appendixes 31 5 Hz 1 1 octave filter 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter...
Страница 265: ...265 SVAN 977C User Manual Appendixes 4 00 Hz 1 1 octave filter 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter...
Страница 311: ...311 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 90 deg 0 deg 90 deg 0 deg 400 Hz 500 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 312: ...312 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 90 deg 0 deg 90 deg 0 deg 1000 Hz 1250 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 313: ...313 SVAN 977C User Manual Appendixes 1600 Hz 2000 Hz 90 deg 0 deg 90 deg 0 deg 2240 Hz 2500 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 314: ...314 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 90 deg 0 deg 90 deg 0 deg 3550 Hz 4000 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 315: ...315 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 90 deg 0 deg 90 deg 0 deg 5600 Hz 6300 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 316: ...316 SVAN 977C User Manual Appendixes 7100 Hz 8000 Hz 90 deg 0 deg 90 deg 0 deg 8500 Hz 9000 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...