123
SVAN 977C User Manual
You can switch on or off consideration of the
ambient noise for the next measurements in
the
Enable
position.
►
Note: The table with ambient noise distortions can be updated also from the Measurement point
screen before the first measurement
– see next Chapter.
13.4.6 STIPA measurements
When
STIPA Settings
are made, you can start measurements.
1. To start measurement, just press the
<Start>
key being in any screen.
Screen with the measurement progress bar and STI index calculated with
considered ambient noise, measured STI index, STI qualification band,
LAeq and LAS results will appear.
The measurement lasts 15 seconds during which the displayed results
are updated every 1 second.
Note: Ambient noise distortions will be considered in accordance with the values set in the
Ambient noise table
– see Chapter
Note: Pressing the <Stop> key during the measurement will break it without recording the
measurement results.
2. After measurement stop the
Measurement point
screen appears.
This screen presents for the measurement point (in the presented case
–
point 1):
•
averaged STI (or CIS) index (
Avg.
) with considered ambient noise
and spread (
Δ
) for the measurement point,
•
STI
(or
CIS
) results with considered ambient noise, STI qualification
band and error indication for all performed measurements (
#1
,
#2
,
..),
•
LAeq
and
LCeq
results for the selected measurement (
# 1
,
# 2
, ..).
3. Press the
<Start>
key if you wish to
perform another measurement.
You can view the modulation function
coefficients for seven octave bands
and two modulation frequencies (f1
and f2) pressing the
<View>
soft-key.
<ENT>
You
can
delete
the
selected
measurement pressing the
<Delete>
soft-key.
You can exclude/include the selected
measurement from the averaging by
pressing the
◄/► key together with
<Alt>
.
<Alt/
►
>
Содержание Shawcity SVAN 977C
Страница 231: ...231 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 1000 Hz 1250 Hz 1600 Hz 2000 Hz...
Страница 232: ...232 SVAN 977C User Manual Appendixes 2240 Hz 2500 Hz 2800 Hz 3150 Hz 3550 Hz 4000 Hz...
Страница 233: ...233 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 5600 Hz 6300 Hz 7100 Hz 8000 Hz...
Страница 234: ...234 SVAN 977C User Manual Appendixes 8500 Hz 9000 Hz 9500 Hz 10000 Hz 10600 Hz 11200 Hz...
Страница 242: ...242 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 243: ...243 SVAN 977C User Manual Appendixes 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 244: ...244 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 245: ...245 SVAN 977C User Manual Appendixes 5600 Hz 6300 Hz 7100 Hz 8000 Hz 8500 Hz 9000 Hz...
Страница 246: ...246 SVAN 977C User Manual Appendixes 9500 Hz 10000 Hz 10600 Hz 11200 Hz 11800 Hz 12500 Hz...
Страница 250: ...250 SVAN 977C User Manual Appendixes...
Страница 262: ...262 SVAN 977C User Manual Appendixes 2 0 kHz 1 1 octave filter 1 0 kHz 1 1 octave filter 500 Hz 1 1 octave filter...
Страница 263: ...263 SVAN 977C User Manual Appendixes 250 Hz 1 1 octave filter 125 Hz 1 1 octave filter 63 0 Hz 1 1 octave filter...
Страница 264: ...264 SVAN 977C User Manual Appendixes 31 5 Hz 1 1 octave filter 16 0 Hz 1 1 octave filter 8 00 Hz 1 1 octave filter...
Страница 265: ...265 SVAN 977C User Manual Appendixes 4 00 Hz 1 1 octave filter 2 00 Hz 1 1 octave filter 1 00 Hz 1 1 octave filter...
Страница 311: ...311 SVAN 977C User Manual Appendixes 250 Hz 315 Hz 90 deg 0 deg 90 deg 0 deg 400 Hz 500 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 312: ...312 SVAN 977C User Manual Appendixes 630 Hz 800 Hz 90 deg 0 deg 90 deg 0 deg 1000 Hz 1250 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 313: ...313 SVAN 977C User Manual Appendixes 1600 Hz 2000 Hz 90 deg 0 deg 90 deg 0 deg 2240 Hz 2500 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 314: ...314 SVAN 977C User Manual Appendixes 2800 Hz 3150 Hz 90 deg 0 deg 90 deg 0 deg 3550 Hz 4000 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 315: ...315 SVAN 977C User Manual Appendixes 4500 Hz 5000 Hz 90 deg 0 deg 90 deg 0 deg 5600 Hz 6300 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 316: ...316 SVAN 977C User Manual Appendixes 7100 Hz 8000 Hz 90 deg 0 deg 90 deg 0 deg 8500 Hz 9000 Hz 90 deg 0 deg 90 deg 0 deg...
Страница 336: ...336 SVAN 977C User Manual Appendixes O ring gasket locations inside...