LAMBDA DG-4 & 5 PLUS/USB OPERATION MANUAL – REV.
iv
filters are accessible via separate panels on the side of the unit. Bulb replacement and
filter installation/removal are covered in separate sections of the manual and should only
be attempted with the power cord disconnected.
To prevent fire or shock hazard do not expose the unit to rain or moisture.
Electromagnetic Interference
To comply with FDA and CE electromagnetic immunity and interference standards; and to
reduce the electromagnetic coupling between this and other equipment in your lab always
use the type and length of interconnect cables provided with the unit for the interconnection
of one or more filter wheels and/or shutters, host computer via serial RS-232 or parallel
interface, or other equipment via TTL connections (see the TECHNICAL SPECIFICATIONS
appendix for more details).
Operational
Operate only in a location where there is a free flow of fresh air on all sides.
NEVER ALLOW THE FREE FLOW OF AIR TO BE RESTRICTED.
Do not operate the lamp in an orientation where the output of the CERMAX lamp faces
within 45
°
of straight upward. To do so may cause arc instability and the possibility of
damage to the front window of the CERMAX lamp. Note that the lamp output is towards
the heat sink mounted on the side of the cabinet.
Avoiding Physical Injury while Powered Up and Emitting Light
WARNING: DO NOT LOOK DIRECTLY INTO THE LIGHT GUIDE! The output
of the light aperture or the light guide should be directed into the microscope using the
appropriate adapters, directed away from anyone’s eyes, and not directed toward any
reflective surface.
INFRARED RADIATION: The infrared radiation (and ultraviolet radiation)
generated by this lamp can cause significant skin burns and eye damage.
THE SMARTSHUTTER IS NOT A SAFETY SHUTTER: The Lambda DG-
4PLUS/USB and DG-5PLUS/USB equipped with a Sutter Instrument
Smart
Shutter. The
SmartShutter is not intended to be a ‘safety shutter.’ A safety shutter usually closes
automatically in the event of a power failure and is designed with the primary goal of
ensuring that it will not allow any unintended exposure. For laser safety applications, a
shutter is normally designed so that no single component failure allows an unintended
exposure to the laser beam. The SmartShutter is intended for use in the controlling of
light in scientific and industrial applications. The SmartShutter was designed for high
performance and durability, but without certain features that would be desirable in a
safety shutter application.
EXPLOSION: High internal pressure exists in any xenon arc lamp.
Содержание Lambda DG-4Plus/USB
Страница 3: ......
Страница 4: ......
Страница 16: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 4...
Страница 24: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 12 Figure 2 7 Lambda DG 4 left side...
Страница 46: ......
Страница 69: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 57 This page intentionally left blank...
Страница 70: ......
Страница 76: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 64 This page intentionally blank...
Страница 78: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 66 This page intentionally blank...
Страница 94: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 82 NOTES...
Страница 98: ...LAMBDA DG 4 5 PLUS USB OPERATION MANUAL REV 1 01B3 20190607 86 NOTES...