LAMBDA 10-3 OPERATION MANUAL – REV. 3.03 (20110829)
37
6.
Save and load configurations
7.
Set power-up default settings (in Lambda 10-3 Firmware Rev. 1.21 and later)
4.6
4.6
4.6
4.6
Selecting the Active Filter Wheel for Manual Control (MODE 1)
Selecting the Active Filter Wheel for Manual Control (MODE 1)
Selecting the Active Filter Wheel for Manual Control (MODE 1)
Selecting the Active Filter Wheel for Manual Control (MODE 1)
When pressing keys 1 through 0 on the keypad while in Local mode, the currently active
filter wheel moves to the corresponding filter position. By default, when the Lambda 10-3 is
first powered on, Wheel A is the active filter wheel. The active filter wheel can be changed by
pressing MODE, then 1, followed by 1 for Wheel A, 2 for Wheel B, or 3 for Wheel C.
4.7
4.7
4.7
4.7
Selecting the Speed for the Active Filter Wheel (MODE 2)
Selecting the Speed for the Active Filter Wheel (MODE 2)
Selecting the Speed for the Active Filter Wheel (MODE 2)
Selecting the Speed for the Active Filter Wheel (MODE 2)
The speed of the currently active filter wheel is set by, while at the Status screen, pressing
MODE, pressing 2, followed by one of eight speeds (keys 0 through 7). The following screen is
displayed after pressing MODE 2 from the Status screen or pressing 2 from MODE screen.
Speed 0 is the fastest speed and 7 is the slowest. Entering the selected speed under computer
control is covered in a separate section.
The switching times (in milliseconds) for each of the 8 speeds is given in Table 4-1. Note that
a 5-filter position move is the longest move on the 10-position filter wheel, since the
controller will always rotate the correct direction to make the fastest change of position. For
a given speed, the time for a move to a non-adjacent filter will take about 1.8, 2.5, 3.3 or 4
times as long as a move to an adjacent filter, depending on the number of filters to be
skipped. The highest speed that can be used will depend on the weight of the filters mounted
in the wheel. In a 25mm, 10-position filter wheel equipped with two typical filters and eight
blocking disks, adjacent filters can be changed in less than 40 milliseconds with a speed
setting of 1. A correspondingly-equipped 32mm wheel may require a lower speed.
SELECT ACTIVE WHEEL(MUST BE CONNECTED):
1-WHEEL A 2-WHEEL B 3-WHEEL C
PLEASE ENTER THE NEW SPEED VALUE BETWEEN
0-7 .....
Содержание Lambda 10-3
Страница 12: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 xii This page intentionally blank ...
Страница 46: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 34 This page intentionally blank ...
Страница 72: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 60 This page intentionally blank ...
Страница 84: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 72 This page intentionally blank ...
Страница 96: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 84 This page intentionally blank ...
Страница 102: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 90 This page intentionally blank ...
Страница 106: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 94 This page intentionally blank ...
Страница 108: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 96 This page intentionally blank ...
Страница 112: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 100 This page intentionally blank ...
Страница 128: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 116 This page intentionally blank ...
Страница 132: ...LAMBDA 10 3 OPERATION MANUAL REV 3 03 20110829 120 NOTES NOTES NOTES NOTES ...