2.3 Assembly of electrode and housing
2.3.1 Cable set-up
a. Make sure to remove the conductive rubber or aluminum-foil layer between the electrode
signal wire and the coaxial shield.
b. Extend the cable to the transmitter without any joint except specific junction box. Connect
the transparent coaxial inner directly to the Glass terminal on the back of transmitter, and
metal connect coaxial shield to Ref
.
terminal.
2.3.2 Assembly of immersive electrode holder and junction box 8-09-5+ PP-100A (Optional)
W/O a temp.
probe
W/ NTC30K W/ PT1000
W/O a temp.
probe, W/ a SG
probe
PP-100A
PP-100AN
PP-100AP
PP-100AE
Note: 1.
PP-xxxA series is applied to 2-wire distribution only.
2. PP-xxxAE series is applied to 3-wire distribution only.
Set-up diagram of coaxial cable:
Follow the left set-up illustration. The black
conductive rubber covering on the coaxial inner
should be removed.
Transparent coaxial inner (remove the conductive rubber)
Metal coaxial shield
A---- Upper cover of round junction box
F---- O-ring
B---- O-ring
G---- PP Electrode Protective Housing
C---- Cable fixing gland MG16A
H---- Electrode (Sensor
)
D---- Lower cover of round junction box
I---- Rubber electrode holder
E---- Cable fixing point MG16A
J---- PP pipe protective cover
PP protective holder:
3