3.4 Installation of accessorial transmitter PH-300T
(Optional)
Accessorial pH/ORP transmitter, PH-300T, is mainly installed on the electrode protective pipe,
but it also can be applied to wall mounting or pipe mounting. For long distance transmission
(100m), for example, if the distance between a transmitter and an electrode overweighs 30
meter long, it is recommended that accessorial transmitter PH-300T should be used to avoid
the attenuation of electrode signal which also leads to convenience of onsite observation,
measurement, and calibration.
Illustration of wall mounting type
1.
Combine the sun shield and PH-300T transmitter by round stainless steel screws 4x15
2.
Fix item 1 combination on wall by self-tapping screws
Illustration of pipe mounting type
1.
Fix immersive holder (B) into U-shaped clamp (A)
2.
Combine the item 1. combination with PH-300T (C) by Round stainless steel screws 4
×
15
A------Sun shield of PH-300T
B------PH-300T transmitter
C------Round stainless steel screws 4x15
D------Self-tapping screws (defined by customer)
A------U-shaped clamp
B------Immersive holder
C------PH-300T transmitter
D------Round stainless steel screws 4
×
15
16