6.2 Analog output setting and RELAY test
1. Under measurement mode, an extended press to
and release until it shows 4mA on
display. It goes analog output of 4mA at the moment.
2. Press
and it shows 20mA on display. It goes analog output of 20mA at the moment.
3. Press
to enter H1 RELAY test and there will show on sign & H1 sign on display. Hl
RELAY is in action and Hi indicator is lighting at the moment.
4. Press
and it will show oFF sign & H1 sign on display. H1 RELAY ceases working and
Hi indicator is being turned off at the moment.
5. Press
to enter L1 RELAY test and there will show on & L1 signs on display. L1
RELAY is in action and Lo indicator is lighting at the moment.
30