8.5 Motor
Enter setup of motor. Select to turn on or turn off the motor operation(can also be set in the setting
mode). Under measurement mode, the motor must be kept running. If the motor stop running, it will
affect the correctness of chlorine measurement. In addition, when the water level detection function is
turned on, under measurement mode if electrode circuit open or water outage(the water level drop to
the bottom of the silver electrode) happens, or if the measurement is kept constantly at near zero, then
the instrument will judge it and stop the motor automatically after a certain of time. It prevents
residual chlorine electrode continuously rub against without water sample and results in damage. It
also shows “M-OFF” warning symbol on the top right corner of the display and goes into HOLD
status. When turning off the water detection function, under measurement mode, regardless of the
electrode circuit open, water outage, or constant zero measurement, the motor will keep running.
Enter
“
Return
”
Setup
Press to confirm it.
Press to confirm it.
Press to confirm it.
Press or to
activate the motor or not.
Press or to turn on
or turn off water level detection
function for auto motor stop.