5. Operation
5.1 Measurement mode:
After all electrical connections are finished and tested, connect the instrument to the power
supply and turn it on. The transmitter will automatically entering measurement mode with the
factory default settings or the last settings from user.
For first operation or long period motor suspension, it is necessary to let the instrument
continuously operate at least one hour before execution when powering on. It makes the electrode
be fully polished by the ceramic beads to ensure the accuracy and stability of the residual chlorine
measurement reading.
5.2 Set-up menu:
Please refer to the set-up instructions in Chapter 7. Press and simultaneously to
enter into set-up menu, and press to
go press to back to measurement
mode.
5.3 Calibration menu:
Please refer to the calibration instructions in Chapter 8. Press and
simultaneously to enter into calibration menu, and press to go back to measurement
mode.
5.4 Shortcuts:
1. In the measurement mode, if selecting MTC for temperature compensation mode, you may press
and to adjust MTC temperature value.
2. Under measurement mode, press continuously for 2 seconds to see the Logbook function
directly. Press key to back to measurement mode.
3. Under measurement mode, press continuously for 2 seconds to switch the display mode
from text mode, trace mode, and real-time chart display mode.
5.5 Default value:
5.5.1 Setting default value:
Measurement display mode: Normal mode(Text & numeral display)
Span calibration: None
Polar. Volt: Initial volt = -100 mV, Gain= -10mV/
μ
A
Polarogram: Starting volt = 0 mV, Finishing volt = -500 mV, Scanning speed = -50 mV/sec,
Max. current display value = 10
μ
A, Min. current display value = -5
μ
A
pH function: OFF
Temperature compensation: ATC
Motor: Auto, Detect ON
Relay 1: Free Cl
2
, High point alarm: AUTO, SP1= 1.00 mg/l, Hys.=0.20 mg/l
Relay 2: Free Cl
2
,
Low point alarm: AUTO, SP2 = 0.30mg/l, Hys.= 0.20 mg/l
Cln/Auto Zero: Clean, OFF
Analog current output: Free Cl
2
, 4~20 mA, 0.00~3.00mg/l
Date & Time: 2013/1/1 00:00:00
RS-485: RTU, 19200, Even, 1, ID: 001
Digital filter: 20
Backlight setting: OFF
Contrast: 0
Logbook: None
Auto return: Auto, 3 minutes
Code set-up: Off
18