1.2 Product model and spare part order number
Model
Sensor Code
Power Code
CT-6110 -POL
POL
:
Polarographic
-03
01: Single-phase, AC 220V(-15~+10%)/60Hz
02: Single-phase, AC 220V(-15~+10%)/50Hz
03: Single-phase, AC 110V(-15~+10%)/60Hz
04: Single-phase, AC 110V(-15~+10%)/50Hz
Standard accessories:
1.
Transmitter
2.
Sensor assembly(Including Cl
2
electrode, temperature probe, motor & driving device)
3.
Flow-through chamber
4.
Measuring cup
5.
Ceramic beads pack x 2
6.
Inlet/Outlet couplings
7.
Wall mounting frame
8.
Transmitter mounting plate
9.
Acrylic cover
10.
Cable gland (If the original glands are not enough for wiring, the additional cable gland is
available for installation at back cover)
If necessary, please order the spare parts as follows:
1.
Free Cl
2
electrode: 8-CL-01
2.
Ceramic beads: 8-40 (Weight about 40g)
3.
pH electrode housing(Optional): 8-41
4.
pH electrode(Optional): Select a suitable pH sensor. When installation, it is necessary to
move the sensor assembly box to the clean position. Open the box upper cover, and install
the pH electrode housing from the box bottom side up. Let the pH cable go through the pH
cable gland of sensor assembly box. Finally, insert the pH electrode into the pH electrode
housing. Please make sure that the pH electrode does exceed the tip of housing.(If the outer
diameter of pH electrode head is too small, the electrode may exceed the tip of the housing.
If so, it is suggested to apply O-ring at the bottom side of pH electrode head.) It will prevent
the pH electrode from damaging by the bottom of the flow-through chamber.
5.
pH buffers (Optional)