Wiring of residual chlorine electrode and temperature probe:
Sensor box
Transmitter
Cl
2
sensor
wiring
Lower level of 6PIN
terminal block for Cl
2
sensor
and temperature probe
Upper level of 6PIN terminal
block for signal cable to
transmitter’s terminal block
Rear terminal
block
Lower end of
rotary joint
○
1
Brown(Thick wire)
○
1
Brown
(11) CE (silver)
Upper end of
rotary joint
○
2
Red(Thick wire)
○
2
Red
(12) WE(gold)
Temperature
probe
○
3
Black
○
3
Black
(15) SG
○
4
Red
○
4
Yellow
(16) T/P
○
5
Green
○
5
Orange
(17) SG
Installation of cable gland:
If the original glands are not enough for wiring, the additional cable gland is available for
installation.
1.
Remove the waterproof plug(A) from the back cover.
2.
Set the rubber gasket(C) in the cable gland (B), and screw into the back cover.
3.
Use a fixing nut(D) from the inside of the back cover to fix up the cable gland(B).
○
6
○
5
○
4
○
3
○
2
○
1
○
6
○
5
○
4
○
3
○
2
○
1
9