Maintenance Procedures
Chapter 4
53
Micro Switch Located Under the Mattress Support
1. Raise the bed to the high position and lock casters.
2. Raise Knee Gatch to maximum height.
3. Manually lift and fold the foot section back towards the head end of the bed (see figure 4.1
in Appendix B).
4. Remove the two screws (
A
) holding the micro switch support (
B
) to the mobile frame right
rail. Remove support.
5. Disconnect the two wires (
C
) from the micro switch (
D
). Note the wire connecting positions.
6. Remove the two screws (
E
) holding the micro switch to its support.
7. Reverse the above steps to install replacement micro switch.
8. Test CPR mechanism for proper operation before returning bed to service.
Figure 4.17A
A
C
E
B
BLACK WIRE
RED WIRE
D
MICRO
-
SWITCH CONNECTIONS
BLACK WIRE
BLACK WIRE