Maintenance Procedures
Chapter 4
37
12. Remove the 12 screws (
A
,
B
, fig. 4.7A, page 32
)
holding the foot end casing cover (
C
, fig.
4.7A, page 32) and the two IV pole holders (
E
, fig. 4.7A, page 32) to the foot end casing.
13. Lift up and hold cover while disconnecting from the PC Board the On/Off switch cable (
F
, fig.
4.7A, page 32) and the PC Board to foot end casing connector cable (
G
, fig. 4.7A, page 32).
Note carefully the connecting positions of the cable connectors to the PC Board connectors
(On/Off switch: cable connector green wire facing pin 11; foot end casing connector cable:
connector black wire facing pin 1). Lay cover aside.
14. Disconnect the night light cable connector (
A
) from the PC Board power cord connector (
B
).
Note the connecting position of the connectors, white wire facing white wire, black facing
black.
15. Cut the cable tie holding cables together.
16. Remove the 3 screws (
C
) holding the rail cover (
D
) to the left rail and slowly remove rail
cover from its location. The night light wiring will follow.
17. Make necessary tests to detect faulty component.
18. Reverse steps to install night light assembly with replacement component.
19. Test night light for proper operation before returning bed to service.
Figure 4.11
D
C
B
A
FOOT END
CASING
NIGHT LIGHT