Chapter 4
The
GO
BED Maintenance Manual
28
4.3 MEMBRANE REPLACEMENT (FOOT BOARD CONTROL PANEL )
N
N
O
O
T
T
E
E
The membrane comes with the control panel plate, already affixed to it.
1. Raise the bed to the high position and lock casters.
2. Unplug the power cord from the wall receptacle.
3. Remove the head board and lay it on a workbench
4. Remove the 7 screws (
A
) securing the support cover (
B
) to the inner face of the foot
board and remove cover.
5. Remove the screw (
C
) holding the membrane support to the foot board.
6. Disconnect the defective membrane cable (
E
) from the board connector (
D
).
7. Remove the control panel plate (
F
).
8. Reverse the above steps to install the replacement membrane.
N
N
O
O
T
T
E
E
Be sure to connect the new membrane cable correctly. The side of the membrane connector
bearing the inscription "1" must face the black wire of the foot board connector.
9.
Test all the foot board control panel functions for proper operation before returning bed
to service.
A
B
C
D
E
F
Figure 4.3