Maintenance Procedures
Chapter 4
25
5. Manually lift and fold the foot section back towards the head end of the bed (see figure 4.1
in Appendix B).
6. Remove the four bolts / lock washers (
A
) holding the side-rail assembly to the foot section.
N
N
O
O
T
T
E
E
Apply medium strength tread locker (blue) on the bolt treads, before replacing the four bolts.
7. Lift up assembly to disengage from anchor point (
B
) and remove assembly
8. Reverse the above steps to install the replacement foot side-rail assembly.
9. Test foot side-rail for proper operation before returning the bed to service.
Head Side-Rail Assembly Replacement
N
N
O
O
T
T
E
E
If the bed being maintain is equipped with the nurse call option, be sure to mention it to the
Service representative when ordering a new head side-rail assembly.
1. Raise the bed to the high position and lock casters.
2. Raise head side-rail to be replaced and raise Fowler to high position.
3. Unplug the power cord from the wall receptacle.
4. Unscrew the lock ring of the side-rail cable plug (
J,
fig. 4.1C, page 25) and remove cable
from the bed receptacle.
5. If the bed being repaired
is not
equipped with the optional nurse call function, proceed with
step 10.
If the bed being repaired
is
equipped with the optional nurse call function, and its version is prior
to version 03 (bed with serial number
[
C02025), proceed with step 6. For version 03 (bed with
serial number
/
C02026) and higher of the
GO
BED (equipped with the optional nurse call
function, proceed with step 10.
6. Remove the fourteen screws (
C
) holding the head end casing (
D
) cover and remove the
cover (
E
).
7. Identify the right (
F
) or left (
G
) nurse call cable wires in the head end casing, cut the cable
tie if necessary.
8. Loosen the two screws holding the two groups of red and black wires on the multiple
connector (
H
), remove both groups of wires and segregate from them the red and black
wires of the nurse call cable involved in the procedure.
AU MUR D'HOPITAL
TO HOSPITAL WALL
C
C
E
D
F
H
G
Figure 4.1B