www.stryker.com
3006-209-001 REV B
73
EMC information
(Continued)
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the 3002
S3
MedSurg bed is used
exceeds the applicable RF compliance level above, the 3002
S3
MedSurg bed should be observed to verify normal
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the 3002
S3
MedSurg bed.
b
Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m.