www.stryker.com
3006-209-001 REV B
55
Operation
Setting the
i
Bed Awareness nurse call alarm
S3
sends a signal through the nurse call system when a parameter condition is compromised.
To set up an
i
Bed
Awareness alarm through the nurse call system:
1.
Press the
Menu
button (A). See
2.
Scroll to
Advanced options
.
3.
Press the
Enter
button (F). See
.
4.
Scroll to
Status to nurse call
.
5.
Press the
Enter
button (F). See
.
6.
Select
On
.
7.
Press the
Enter
button (F). See
.
Note:
A confirmation notification indicates that you set the alarm tone.
i
Bed Awareness (option)
C
A
D
F
E
G
B
Note:
For a description of the menu functions, see
A
i
Bed
Arms/disarms
i
Bed
Awareness
B
Menu
Accesses menu functions
C
Display
Displays menu functions
D
Up arrow
Scroll up through menu functions
E
Down arrow
Scroll down through menu functions
F
Exit
Exits from menu functions or cancels
operation
G
Enter
Selects menu functions or saves
operation