www.stryker.com
3006-209-001 REV B
43
Operation
Footboard LED indicators (Continued)
Indicator
LED illuminates amber
C
Patient control Fowler lock
The patient control Fowler lock is on
D
Patient control gatch lock
The patient control gatch lock is on
E
Patient control bed height lock
The patient control bed height lock is on
F
Zone 1
Bed exit is on and zone 1 is active. The LED will
blink amber if a bed exit event occurs.
G
Zone 2
Bed exit is on and zone 2 is active. The LED will
blink amber if a bed exit event occurs.
H
Zone 3
Bed exit is on and zone 3 is active. The LED will
blink amber if a bed exit event occurs.
A
D
E
F
G
H
C
B
Figure 27: Footboard LED indicators
Indicator
LED illuminates amber
A
Bed motion lock
Bed motion lock is activated or when the patient
control (Fowler, gatch, bed up/down) lock buttons
are activated
B
Fowler 30°+
Fowler 30°+ is locked. The LED will blink amber if:
•
i
Bed
Awareness system is on
•
Fowler 30°+ is being monitored and the Fowler
goes below 30 degrees
•
Fowler 30°+ is turned off
C
Low height
Bed is in low height. The LED will blink amber if:
•
i
Bed
Awareness system is on
•
Low height is being monitored
•
Bed is not in low height
D
Brake
Brake is set, and will blink amber if the brake is not
set
E
Bed exit (optional)
Bed exit is armed. The LED will blink amber if:
•
Bed exit is turned off while the
i
Bed
Awareness
system is turned on
•
Bed exit alarms while monitored by the
i
Bed
Awareness system
F
Bed zero (optional
i
Bed
Awareness)
Bed zero is successful