52
3006-209-001 REV B
www.stryker.com
Operation
Changing the patient weight (Continued)
3.
Press and hold the
Enter
button (F) (see
) (
) until
Release Button
appears (
).
Figure 48: Hold to change patient weight
Figure 49: Release Button
4.
Release the
Enter
button.
Note:
Do not touch the product when configuring patient weight (
Figure 50: Do not touch bed
5.
When the system is ready to change patient weight, press the
Up arrow
button (C) or
Down arrow
button (D) to
change the displayed weight (see
).
6.
After changing the weight, press the
Enter
button (F) (see
) and the message
Patient
Weight Changed
will display.
To cancel the request, press the
Exit
button (E) (see
) and the message
Operation Canceled
will display.
Changing the scale units
You can change the measuring unit to pounds (lb) or kilograms (kg) on your display.
Note:
The default scale unit is pounds (lb).
To change the displayed scaled units:
1.
Press the
Menu
button (A). See
2.
Scroll to
Scale Units
.
Содержание S3 MEDSURG BED
Страница 2: ...sample text ...
Страница 67: ...www stryker com 3006 209 001 REV B 63 Operation iBed Locator option Continued A Figure 59 iBed Locator ...
Страница 72: ...68 3006 209 001 REV B www stryker com Preventive maintenance Product serial number Completed by Date ...