50
3006-209-001 REV B
www.stryker.com
Operation
Measuring weight gain or loss (Continued)
4.
Release the
Enter
button.
Note:
Do not touch the product when you measure weight gain or loss (
).
Figure 42: Do not touch bed
A confirmation notification indicates that enabling
Gain/Loss
was successful.
The base measurement is the initial patient weight registered when you enable
Gain/Loss
. The weight difference
between the base weight and the weight gained or lost is displayed on the bottom right hand corner (
Figure 43: Base and gain/loss weight measurement
Note:
If the weight gained or lost exceeds 99.9 lb.,
Error –.-
appears on the display (
Figure 44: Gain/loss error
To disable gain or loss:
1. Press the
Menu
button (A). See
.
2. Scroll to
Gain/Loss
.
3. Press and hold the
Enter
button (F). See
Changing equipment
Change equipment allows you to add or remove equipment or devices from the product without affecting the patient
weight.
To change equipment:
1.
Press the
Menu
button (A). See