3006-209-001 REV B
13
Introduction
System requirements and recommendations for
i
Bed Wireless (option)
(Continued)
Customer WLAN environment
Required
Channel utilization
Consistently less than 30%
Recommended
Signal strength range (minimum)
2.4 GHz: -67dBm +0/-8dBm
5 GHz: -67dBm +0/-8dBm
Yes
Minimum SNR
Minimum 20dB
Yes
Priority queuing
Prioritized over best effort traffic
Recommended
Client exclusion
Disabled
Recommended
Client load balancing
Disabled
Recommended
Max number of SSIDs
5
Recommended
Authentication timeouts
Add session timeout of at least 24
hours
Recommended
Note:
A transmit power asymmetry problem may arise at the edges of virtual cell coverage if an APs transmit power is
higher than the Stryker Wireless Client device (~6 mW 2.4 GHz or 12 mW 5 GHz). The received signal strength indicator
(RSSI) of the Stryker
i
Bed
Wireless Client on the AP must be verified. The device should never drop below an RSSI of
-75 dBm on the AP.