To Connect a Ground Pad (monopolar only):
1. Plug the generator end of the ground pad cable into the designated port
on the generator. Use the ground pad port symbol as reference and
match the white band on the port to the white cable strain relief.
2. Raise the connector latch at the other end of the cable, insert the
ground pad tabs into the connector and close the latch to secure the
ground pad.
WARNINGS:
• Place the ground pad with its entire surface securely on the patient’s
body in proximity to the treatment site.
• Properly affix the ground pad to the patient. Alcohol may be used to
clean the site and eliminate poor adhesion caused by a wet surface
or the use of lotion. Verify proper contact exists between the patient
and the ground pad whenever the patient is repositioned after the initial
application of the ground pad.
• ALWAYS make sure a safe contact exists between the ground pad and
the patient. An auditory alarm will not occur if safe contact between
the ground pad (neutral electrode) and the patient is not established or
maintained.
• Properly locate and orient the ground pad on the patient to control the
size and shape of the lesion during its creation.
NOTE:
ALWAYS use and properly connect a ground pad to the generator
for monopolar electrode configurations. The generator will not energize
the monopolar electrode if a ground pad is not adequately connected and
applied to the patient. If a ground pad is not connected or the connection
is interrupted, an error message will appear on the display.
3. Place the pad portion of the ground pad onto the patient, in proximity to
the treatment site (see figure 6). The entire pad should adhere to the
patient's body.
Figure 6 Pad Placement
To Connect a Printer (optional):
NOTE:
As a minimum requirement, any printer used with the generator
must be PCL3 compatible. Printers using versions of PCL greater than 3
may also work.
1. If using a printer, connect the printer cable to the parallel printer cable
port on the generator and the appropriate printer receptacle.
2. Connect facility power to the printer.
To Connect a RF Hand Controller (optional):
1. If a RF Hand Controller is used (see Appendix A -
RF Hand Controller
Instructions
section), align the dot on the cable connector end with the
dot on the generator port and connect the cable to the generator.
2. Use the RF Hand Controller port symbol as a reference and match the
color-coded gray band on the port with the gray cable strain relief to
ensure the installation is correct.