0406-900-702 Rev-E
EN
www.stryker.com
53
NOTES:
• At an electrode spacing of 10 mm, two independent lesions are formed,
one around each electrode. The lesions tend to show a slight inclination
to form toward the other electrode as shown in the 10 mm example.
• At an electrode spacing of less than 10 mm, one lesion is formed
between the two electrodes. For more information, see Pino, Hoeft et al
1
.
1
Pino CA, Hoeft MA, Hofsess C, Rathmell JP (2005) Morphologic analysis of bipolar
radio frequency lesions: Implications for treatment of the sacroiliac joint. Regional
Anesthesia and Pain Medicine, Vol. 30, No 4 (July – August), 2005: 335–338.
10 mm
Figure 49 Non-contiguous Lesions 4.4 mm x 7.5 mm (width x length)
5.0 mm x 8.2 mm (width x length)
14 mm
Figure 50 Non-contiguous Lesions 4.7 mm x 8.5 mm (width x length)
5.1 mm x 8.2 mm (width x length)
10 mm
14 mm