0406-900-702 Rev-E
EN
www.stryker.com
45
Specifications
Model:
REF 0406-900-000 RF MultiGen Generator
Size:
12.5 inch [317.5 mm] width, 8 inch [203.2 mm] height, 15 inch [381 mm] depth
Weight:
18 lb. [8.2 kg]
Equipment Type:
Class 1
Type BF Applied Part
Power Supply:
100 - 120 VAC
50 - 60 Hz, 230 VAC
50 - 60 Hz
Fuse Type and Rating
2 x 2.5 A, 250 VAC
, (T) time-lag (slow-blow), 5 x 20 mm, IEC 127
Power Output:
50 watt maximum power into 100 ohm resistive load
Enclosure Protection:
IPX0 ordinary equipment
Protective Earth Ground:
Mode of Operation:
Non-continuous
Duty Cycle:
999 seconds ON, 90 seconds OFF
PRODUCT SAFETY CERTIFICATION STANDARDS:
CSA International
CANADIAN STANDARDS ASSOCIATION (CSA)
CAN/CSA-C22.2 No. 60601-1:08,
Medical Electrical Equipment — Part 1: General
Requirements for Basic Safety and Essential Performance
AMERICAN NATIONAL STANDARDS INSTITUTE (ANSI)/ASSOCIATION FOR THE
ADVANCEMENT OF MEDICAL INSTRUMENTATION (AAMI)
ANSI/AAMI ES60601-1:2005,
Medical Electrical Equipment — Part 1: General
Requirements for Basic Safety and Essential Performance
; Consolidated Reprint
(2009); Amendment 2 (2010)
INTERNATIONAL ELECTROTECHNICAL COMMISSION (IEC)
IEC 60601-1:2005, Ed:3,
Medical Electrical Equipment — Part 1: General
Requirements for Basic Safety and Essential Performance
; Corrigendum 1 (2006);
Corrigendum 2 (2007)
EUROPEAN COMMITTEE FOR ELECTROTECHNICAL STANDARDIZATION
(CENELEC)
EN 60601-1:2006, Ed:3,
Medical Electrical Equipment – Part 1: General Requirements
for Basic Safety and Essential Performance;
IEC Corrigendum 1 (2006); IEC
Corrigendum 2 (2007); CENELEC Corrigendum (2010); CENELEC Amendment A11
(2011)
Display Screen:
5.5 inch x 8 inch liquid crystal diode, wide 160° minimum viewing angle
Impedance:
Measurement range:
Accuracy:
Operating range:
0 ohm - 2000 ohms
below 100 ohms + 30 ohms; above 100 ohms + 10%
35 ohms<stimulation<1800 ohms; 35 ohms<lesion<1800 ohms
Temperature:
RF Procedure Accuracy:
Maximum Set Temperature (Thermal mode):
Maximum Set Temperature (Pulse mode, 2 Hz/ 20 ms, 20 Hz/ 10 ms):
Maximum Set Temperature (Pulse mode, 2 Hz/ 100 ms):
Maximum Set Temperature (Pulse mode, unless specified otherwise):
+4 °C / -2 °C from 37 °C to 95 °C
95 °C
50 °C
60 °C
80 °C
Содержание RF MultiGen 0406-900-000