To Change System Settings: Procedure Default
Settings
From the System Settings screen, touch the PROCEDURE DEFAULT button
to access the Set Procedure Default screen (see figure 41). See
Appendix
F - Factory Default Values
section.
1
2
3
4
SENSORY
MOTOR
LESION
Frequency
50.0 Hz
Width
0.75 ms
Amplitude
3.00 v
Frequency
2.0 Hz
Width
1.00 ms
Amplitude
5.20 v
Temp. Limit (c):
80
Hold Time (Sec):
45
PULSE
Temp. Limit (c):
50
Hold Time (Sec):
30
Pulse Freq. (Hz):
20
Pulse Width (ms):
2
Voltage (v):
5
EDIT
EDIT
EDIT
ACCEPT
CANCEL
APPLY TO THIS TAB ONLY
APPLY TO ALL TABS
PARALLEL
PARALLEL
IDL
Figure 41 Set Procedure Default Screen
A
Electrode Identifier Number
Touch to view the procedure default setting values for a specific
procedure tab selection.
B
Procedure Tab Selection Button
Touch to select all procedure tabs or individual tabs when
modifying default setting values. If APPLY TO ALL TABS is selected,
the values for the current selected tab will be applied to the other
three tabs. If APPLY TO THIS TAB ONLY is selected, each tab may
be edited individually.
C
EDIT Settings Buttons
Touch to edit the setting values for the sensory, motor or lesion
procedures.
D
ACCEPT Button
Touch to accept changes to the procedure default setting values.
E
CANCEL Button
Touch to cancel any changes made to the procedure default
setting values.
F
IDL Button
The intradiscal adapter cable and catheter procedure are no longer
available.
PARALLEL
PARALLEL
SUMMARY
SENSORY
MOTOR
LESION
SET TEMP
80c
HOLD TIME
90s
PLACEMENT
L4 LEFT
Thermal
Pulse
EDIT
EDIT
EDIT
RF OUTPUT
TEMPERATURE
TIME
IMPEDANCE
WATTAGE
TEMPERATURE
TIME
IMPEDANCE
WATTAGE
EDIT
PARALLEL
TEMPERATURE
TIME
IMPEDANCE
WATTAGE
EDIT
TEMPERATURE
TIME
IMPEDANCE
WATTAGE
EDIT
PARALLEL
4
3
2
1
1
Figure 42 Typical Edit Procedure Default Settings Screen
To Edit Procedure Default Settings
From the Set Procedure Default screen, touch the EDIT button to access
an Edit Procedure Default Settings screen (see figure 42). See
To EDIT
Sensory or Motor Stimulation Settings
section or
To Edit RF Output Settings
section to edit the procedure default setting values. To return to the Set
Procedure Default screen, touch the SUMMARY button.
A
B
C
E
D
F