0406-900-702 Rev-E
EN
www.stryker.com
25
To EDIT RF Output Settings (continued)
RF OUTPUT
SET TEMP
42c
HOLD TIME
120s
PLACEMENT
L3 RIGHT
PULSE VALUES
2Hz
20ms
Thermal
Pulse
EDIT
EDIT
EDIT
EDIT
PARALLEL
PARALLEL
PARALLEL
PARALLEL
4
3
2
1
1
SUMMARY
START
1xxx
SENSORY
MOTOR
LESION
EDIT
EDIT
EDIT
EDIT
EDIT
EDIT
37
C
0
S
248
Ω
70
MA
45
v
TEMPERATURE
TIME
IMPEDANCE
CURRENT
VOLTAGE
37
C
0
S
248
Ω
---
MA
---
v
TEMPERATURE
TIME
IMPEDANCE
CURRENT
VOLTAGE
37
C
0
S
248
Ω
---
MA
---
v
TEMPERATURE
TIME
IMPEDANCE
CURRENT
VOLTAGE
37
C
0
S
248
Ω
---
MA
---
v
TEMPERATURE
TIME
IMPEDANCE
CURRENT
VOLTAGE
Figure 20 RF Output Screen in PULSE Mode - INACTIVE
2
Hz
20
ms
ENTER
CANCEL
2/20
100/2
50/4
20/10
2/100 400/0.5
EDIT RF - PULSE
Figure 21 RF Output PULSE Preset Dashboard
WARNING:
When attempting PULSE RF lesion creation,
inadvertent selection of the THERMAL RF button may result in
undesired tissue damage and result in patient injury.
NOTE:
THERMAL and PULSE modes of operation may be performed at
the same time in separate procedure tabs.
A
EDIT PULSE VALUES Button/Window
Available in PULSE mode only, touch to select a preset value and
a PULSE CONTROL option. The selected value appears in the
window.
B
NEW VALUE Window
Displays the selected new value.
C
Preset Value Buttons
Touch to select a preset setting.
D
THERMAL Button
Touch the THERMAL RF mode button to select the THERMAL RF
mode of operation. This mode of operation provides high dose
RF energy to elevate the temperature of the target tissue in a
controlled manner.
E
PULSE Button
Touch the PULSE RF mode button to select the PULSE RF mode
of operation. This mode of operation provides pulsed or bursts
of RF energy to stimulate the target tissue with a minimum
temperature elevation.
A
B
C
D
E