0406-900-702 Rev-E
EN
www.stryker.com
27
To Control RF Output:
To Start Single Electrode(s) - separately or staggered (continued)
PARALLEL
PARALLEL
37
C
0
S
250
Ω
---
w
TEMPERATURE
TIME
IMPEDANCE
WATTAGE
80
C
90
S
310
Ω
70
w
TEMPERATURE
TIME
IMPEDANCE
WATTAGE
EDIT
PARALLEL
37
C
0
S
250
Ω
---
w
TEMPERATURE
TIME
IMPEDANCE
WATTAGE
EDIT
37
C
0
S
250
Ω
---
w
TEMPERATURE
TIME
IMPEDANCE
WATTAGE
EDIT
PARALLEL
4
3
2
1
1
STOP
ALL
SET TEMP
80
c
HOLD TIME
90s
PLACEMENT
L4 LEFT
RF OUTPUT
PARALLEL
PARALLEL
EDIT
PARALLEL
EDIT
EDIT
PARALLEL
4
3
2
1
1
STOP
ALL
SET TEMP
42
c
HOLD TIME
120s
PLACEMENT
L4 LEFT
RF OUTPUT
42
C
120
S
248
Ω
70
MA
45
v
TEMPERATURE
TIME
IMPEDANCE
CURRENT
VOLTAGE
37
C
90
S
248
Ω
---
MA
---
v
TEMPERATURE
TIME
IMPEDANCE
CURRENT
VOLTAGE
37
C
90
S
248
Ω
---
MA
---
v
TEMPERATURE
TIME
IMPEDANCE
CURRENT
VOLTAGE
37
C
90
S
248
Ω
---
MA
---
v
TEMPERATURE
TIME
IMPEDANCE
CURRENT
VOLTAGE
Figure 23 RF Output Screen THERMAL mode - ACTIVE
Figure 24 RF Output Screen PULSE mode - ACTIVE
A
TEMPERATURE Tab Value
• The actual value, displayed in degrees Celsius, is continuously
monitored and represents the temperature at the electrode tip.
• If no electrode is connected, the tab will display “----.”
B
TIME Tab Value
Value displayed represents the HOLD TIME value remaining in
seconds.
C
IMPEDANCE Tab Value
• The value, displayed in ohms [0-2000 Ω], is continuously
monitored and represents tissue impedance measured by the
electrode(s).
•
If no electrodes are connected, the IMPEDANCE tab value will
display “- - -.” If an electrode is connected to any port, but the
value exceeds its limit, the value displayed will be “HIGH” for all
tabs.
• A lesion cannot be created if the impedance value is outside the
range of 35 to 1800 Ω.
D
WATTAGE Tab Value
• Available in the THERMAL mode only, continuously displayed
during active RF energy output and represents the approximate
power output.
• If no electrode is connected, output is not active, or output is
less than 0.1 W, the window will display “----W”.
E
SET TEMP Window
Indicates the SET TEMPerature value in the THERMAL or PULSE
mode.
F
HOLD TIME Window
Indicates the HOLD TIME value in the THERMAL or PULSE mode.
G
CURRENT Tab Value
• Available in the PULSE mode only, the value displayed
represents the current in milliamps.
• If no electrode is connected or output is not active, the window
will display “----mA”.
H
VOLTAGE Tab Value
• Available in the PULSE mode only, the value displayed
represents the voltage in volts.
• If no electrode is connected or output is not active, the window
will display “----V”.
I
STOP ALL Button
Touch to remove the application of RF energy output from all
active electrodes.
A
B
C
D
E
F
G
H
I
Содержание RF MultiGen 0406-900-000
Страница 3: ......