![Stryker FL26E Скачать руководство пользователя страница 37](http://html.mh-extra.com/html/stryker/fl26e/fl26e_maintenance-manual_1385904037.webp)
Maintenance
Procedures
Chapter 4
37
2
nd
micro switch:
When the Fowler is lowered while the bed in the
Auto Contour
position, this
switch tells the control board to start lowering the Knee Gatch.
9. Using a Phillips screwdriver, remove the two screws (G) holding the support to the frame and
remove the support.
10. Remove the cable wires from the switch. Note their location so they will be connected
properly to the new switch.
11. Using a #1 Phillips screwdriver, remove the two screws (H) holding the micro switch to the
support.
12. Reverse the above steps to install the new micro switch.
13. Test the
Auto Contour
positioning for proper operation before returning the bed to service.