![Stryker FL26E Скачать руководство пользователя страница 28](http://html.mh-extra.com/html/stryker/fl26e/fl26e_maintenance-manual_1385904028.webp)
Maintenance Manual
28
3. Lower the siderails.
4. Lift and fold the foot section back toward the head end of the bed.
5. Remove the foot board.
6. Using long nose pliers, remove the Rue ring cotter/washers/clevis pin (A) linking the head
actuator tube to the lower part of the head section lever.
7. Using a Phillips screwdriver, remove the six screws (A, fig. 4.3D, page 33) holding the cover
to the foot casing. Grab both ends of the cover and lift up to remove it.
8. Properly ground yourself (see section 1.4).
9. Using side cutters, clip the cable tie holding the head actuator cable to the other cables.
10. Remove the actuator cable (connector J8) from the control board and pass the cable
connector through the rear casing hole.
11. Remove from the capacitor the two wires (B) connected to it and remove, using a Phillips
screwdriver, the screw (C) holding the capacitor to the foot casing. Dispose of the capacitor.
12. Using a Phillips screwdriver, remove the two screws (D) holding the retaining plate to the
actuator support and remove the plate.
13. Remove the two pivot pins (E) holding the actuator to the support. To facilitate the removal
of the pins, insert a small slotted head screwdriver into the opening at the end of the
actuator and push out the pins.
14. Move the defective actuator toward the centre of the bed to remove it from its location.
15. Reverse the above steps to install the new actuator. Carefully read the following caution
before hooking up the actuator tube to the head section lever arms.
It is of utmost importance
that the course of the new head actuator be adjusted before
hooking up its tube to the lower part of the head section lever. An improper adjustment can
damage the head section structure.
16. To adjust the course of the replacement head actuator, proceed as follows:
A. Make sure that the actuator cable is connected to the control board and connect the
bed power cord.
B. Grab the actuator tube and position its holes horizontally. While holding the tube firmly
to prevent it from rotating, press the Fowler up control for a few seconds, then press
the down control until the actuator stops. This will be the lower limit of the actuator
course.
C. Gently turn the tube in either direction to align the tube holes with those of the head
section lever. Then raise again the Fowler a few inches while holding firmly the tube
and lower it completely.
D. Check the alignment of the holes. If they are not aligned any more, repeat steps
B
and
C
until they are. Once the holes are aligned, install the washers and the clevis pin.
Slightly raise the head section manually to ease the insertion of the clevis pin.
E. Before installing the rue ring cotter to finalize the head actuator replacement, raise and
lower completely the Fowler. Check that the actuator stops working as soon as the
head section reaches the frame.
C
C
A
A
U
U
T
T
I
I
O
O
N
N
Содержание FL26E
Страница 2: ......
Страница 45: ...Appendix A Circuit Diagrams 45 ...
Страница 46: ...Appendix A Circuit Diagrams 46 ...
Страница 47: ...Appendix A Circuit Diagrams 47 ...