Maintenance Manual
12
2.3 PREVENTATIVE MAINTENANCE PROGRAM
When servicing beds, use only identical replacement parts provided by Stryker.
ANNUAL CHECKLIST
⎯
All fasteners secure.
⎯
Inspect for excessive wear the oil-impregnated bronze shoulder spacers found at the bed
hinge points. Replace as needed.
Do not
lubricate these spacers.
⎯
Check the grease present on the actuator screws, lubricate if needed (see section2.2). The
actuator screws need lubrication every two years.
⎯
On both sides of the bed, depress fully down the side of the pedal identified with a red
sticker and ensure that the brakes are applied and the bed immobilized. Toggle the pedal to
neutral and ensure the brakes are released.
⎯
On both sides of the bed, depress fully down the side of the pedal identified with a green
sticker and ensure that the steer wheel is engaged. Toggle the pedal to neutral and ensure
that the steer wheel disengages.
⎯
Siderails move, latch and stow properly.
⎯
All functions of the foot end control panel working properly, including LEDs.
⎯
All siderail controls working properly.
⎯
Ensure that the optional nurse call works properly and that the alarm sounds in the nurse
station.
⎯
Verify the CPR emergency release (optional) using both CPR release handles: raise the
Fowler fully up and, using the CPR handle, lower the Fowler gradually to flat position by
pulling, holding and releasing the handle several times. Ensure the Knee Gatch (if raised)
also starts flattening when the Fowler is completely down. Following the complete lowering
of the Fowler, wait approximately 30 seconds - the time for the Fowler control motor to reset
itself - and verify that the motor has indeed reset itself by raising the Fowler fully up using
the Fowler up control.
⎯
Verify the Fowler, Knee Gatch and Hi-lo movements to ensure that the limit switch
integrated to the four electric actuators is operating properly.
⎯
Auxiliary outlet (option available only with 120V beds) working properly.
⎯
Night light (optional) working properly.
⎯
No cracks in the boards, siderails, caster covers and 5th wheel hood (optional).
⎯
Head end bumpers tightly secured to frame and working properly.
⎯
No rips or cracks in mattress cover.
⎯
Power cord not frayed.
⎯
No cables worn or pinched.
⎯
All electrical connections tight.
⎯
All grounds secure to the frame.
⎯
All casters roll properly. Check caster for cuts, wear, etc.
⎯
Measure current leakage and grounding continuity of the bed and the optional auxiliary
outlet. Check with our Technical Service department for the acceptable values.