Maintenance Manual
26
HEAD SECTION REPLACEMENT
Required Tools:
Ratchet w/1/2" Long Socket
1/2" Wrenches (2)
Side Cutters
Torque Wrench
Procedure:
N
N
O
O
T
T
E
E
Unless otherwise indicated, refer to figure 4.3A on page 24 for the illustration of this procedure
reference points.
Under no circumstances should the CPR emergency release handles be activated during this
procedure or serious injury to people or damage to equipment could occur.
1. Fully raise the bed and apply the brakes.
2. Fully raise the head section and remove the head board.
3. Unplug the bed power cord from the wall outlet and raise the head siderails.
4. Loosen the lock rings (A, fig. 4.1A, page 17) and unplug the siderail control cables.
5. Using a ratchet with a 1/2" long socket, remove the eight thread-rolling bolts (B, fig. 4.1A,
page 17) holding the head siderail assemblies to the head section and remove the siderails.
Support the assembly when removing the last bolts.
N
N
O
O
T
T
E
E
Once removed, the thread-rolling bolts must first be screwed in manually before tightening them
to a 130 lbf in (14.7 N m) torque using a torque wrench.
6. Using side cutters, remove the cable ties holding the cables on each side of the head
section structure as well as the cable ties holding the CPR micro switch cables.
7. Using a 1/2" wrench, remove the nuts/shoulder spacers/bolts (J) holding the head section to
the stabilizers. Lower the stabilizer.
8. Plug in the bed and lower the head section. Push slightly on it to ease its descent. Unplug
then the bed power cord.
9. Using long nose pliers, remove the Rue ring cotter/washers/clevis pin (A, fig. 4.3B, page 27)
holding the lower part of the head section lever to the head actuator tube.
10. Using two 1/2" wrenches, remove the bolt/shoulder spacers/nut (N, fig. 4.3H, page 38)
holding the upper part of the head section lever to the head section.
11. Lay head section assembly upside down on a workbench.
12. Remove the two side mattress retainers and install them on the replacement head section.
13. Remove the CPR mechanism and install it on the replacement head section.
14. Reverse the above steps to install the new head section.
15. Test the CPR mechanism for proper operation before returning the bed to service.
W
W
A
A
R
R
N
N
I
I
N
N
G
G