KNEE GATCH ACTUATOR REPLACEMENT
Required Tools:
Long Nose Pliers
Phillips Screwdriver
Pliers
Side Cutters
Strap
Small Slotted Head Screwdriver
Procedure:
N
N
O
O
T
T
E
E
Unless otherwise indicated, refer to figure 4.3B on page 27 for the illustration of this procedure
reference points.
1. Fully raise the bed and apply the brakes. Flatten the mattress support.
2. Unplug the bed power cord from the wall outlet.
3. Lower the siderails.
4. Lift and fold the foot section back toward the head end of the bed. Secure its position using
a strap.
5. Using long nose pliers, remove the rue ring cotter/washers/clevis pin (F) linking the actuator
tube to the thigh section lever arms.
6. Using a Phillips screwdriver, remove the six screws (A, fig. 4.3D, page 33) holding the cover
to the foot casing. Grab both ends of the cover and lift up to remove it.
7. Properly ground yourself (see section 1.4).
8. Using side cutters, clip the cable tie holding the thigh actuator cable to the other cables.
9. Remove the actuator cable (connector J9) from the control board.
10. Using pliers, squeeze the upper part of the strain-relief bushing (G) and lift up to remove it
from its location.
11. Remove the cable from the bushing and pass the cable connector through the casing hole
previously occupied by the bushing.
12. Using a Phillips screwdriver, remove the two screws (H) holding the retaining plate to the
actuator support and remove the plate.
13. Remove the two pivot pins (I) holding the actuator to the support. To facilitate the removal of
the pins, insert a small slotted head screwdriver into the opening at the end of the actuator
and push out the pins.
14. Move the defective actuator toward the centre of the bed to remove it from its location.
15. Reverse the above steps to install the new actuator. Carefully read the following caution
before hooking up the actuator tube to the thigh section lever arms.
It is of utmost importance
that the course of the new thigh actuator be adjusted before
hooking up its tube to the thigh section lever arms. An improper adjustment can damage the
thigh section structure.
19. To adjust the course of the replacement thigh actuator, proceed as follows:
A. Make sure that the actuator cable is connected to the control board and connect the
bed power cord.
B. Grab the actuator tube and position its holes horizontally. While holding the tube firmly
to prevent it from rotating, press the Knee Gatch up control for a few seconds, then
press the down control until the actuator stops. This will be the lower limit of the
actuator course.
C. Gently turn the tube in either direction to align the tube holes with those of the thigh
section lever arms. Then raise again the Knee Gatch a few inches while holding firmly
the tube and lower it completely.