![Stryker FL26E Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/stryker/fl26e/fl26e_maintenance-manual_1385904010.webp)
Maintenance Manual
10
2.2 LUBRICATION REQUIREMENTS
The only components of the bed needing periodic lubrication are the four actuator screws. They
should be checked every year and lubricated every two years.
The bed uses oil-impregnated shoulder spacers at hinge points.
Do not
lubricate these shoulder
spacers. When shoulder spacers are found worn, replace them.
ACTUATOR SCREW LUBRICATION
Required Tools:
Phillips Screwdriver
1/4" Ratchet (w/6" extension) and 5/16" socket
OG2
Grease
Brush
Bungee
Cord
Procedure:
Head and Thigh Actuators
1. Fully raise the bed and apply the brakes.
2. Fully raise the head section and flatten the thigh section to expose maximum screw threads
on both actuators. Lower the four siderails.
3. Unplug the power cord from the wall outlet.
4. Lift and fold back the foot section toward the head end of the bed. Secure its position using
a bungee cord.
5. Using a Phillips screwdriver, remove the four screws (A) holding the cover plate to the frame
and remove the plate. Remove the night light, if applicable.
6. Using a 1/4" ratchet (w/6" extension) and a 5/16" socket, remove the two screws (B) holding
the dust tube of each actuator.