S
Se
errvviicce
e
P
Prro
otte
eccttiin
ng
g a
ag
ga
aiin
nsstt e
elle
eccttrro
osstta
attiicc d
diisscch
ha
arrg
ge
e ((E
ES
SD
D))
C
CA
AU
UT
TIIO
ON
N
•
Always use ESD protective equipment before you open antistatic bags and service electronic parts.
•
Do not place unprotected circuit boards on the floor.
N
No
otte
e -- Always ship the circuit boards back to Stryker. Use the antistatic bag that the new board was originally shipped in.
The electronic circuits in the product are completely protected from static electricity damage when factory assembled.
Always use adequate static protection when you service the electronic systems of the product. All service personnel must
use static protection whenever they touch wires.
Sample antistatic protection equipment includes:
•
Antistatic wrist strap
•
Grounding plug
•
Test lead with a banana plug on one end and an alligator clip on the other end
Make sure that you follow the ESD manufacturer’s instructions for appropriate protection against static discharge.
C
Ca
asstte
err rre
ep
plla
acce
em
me
en
ntt,, ffiixxe
ed
d b
ba
asse
e
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Floor jack
•
3/4” socket
•
(2) Bungee cords (or equivalent)
•
Stryker special key (19-0803Z)
P
Prro
occe
ed
du
urre
e::
1. Raise all siderails to the highest position.
2. Lift the base hood.
3. Separate the V
Ve
ellccrro
o® fasteners that secure the base hood to the base.
4. Using bungee cords, support the base hood.
5. Using a floor jack, raise the end of the product near the caster that needs repair approximately nine inches off the floor.
6. Using the Stryker special key (19-0803Z) and a 3/4” socket, remove the locknut and washer (A) that secure the caster
(B) to the base (Figure 5). Discard the caster. Save the locknut and washer.
1900-009-002 Rev AA.6
13
EN