Pay attention to the position of the cable ties and the slack of the cable between each cable tie.
•
Make sure that the load cell has a one inch radius curve at the first point where it is secured when you install new
cables ties on the load cell cable. Leave a small amount of slack in the cable while you install the remainder of the
cable ties. This allows for product movement.
9. Using a 9/16” combination wrench, remove the bolt (D), washer (C), and spacer (B) that secure the fixed socket (A) to
the accessory bracket (Figure 18). Save the bolt, washer, and spacer.
A
B
C
D
F
Fiig
gu
urre
e 1
18
8 –– F
Fiixxe
ed
d sso
occkke
ett
10. Using a small regular screwdriver, remove the load cell cable from the control board.
11. Pass the connector through the opening in the lower cover plate. Remove and discard the load cell.
12. Reverse steps to reinstall.
N
No
otte
e -- Use a 0.036 in. shim to create a gap between the new load cell and the nylon support.
13. Calibrate the scale system. See
14. Verify proper operation before you return the product to service.
A
An
ng
glle
e sse
en
nsso
orr rre
ep
plla
acce
em
me
en
ntt,, sscca
alle
e o
op
pttiio
on
n
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
#2 Phillips screwdriver
•
3/8” combination wrench
P
Prro
occe
ed
du
urre
e::
1. Raise the endrail at the foot end to the highest position and lower the siderails all the way down.
2. Lift the foot section and fold it toward the head end of the product.
3. Using a #2 Phillips screwdriver, remove the seven screws that secure the foot cover plate to the litter frame. Remove
and save the cover plate (Figure 16). Save the screws.
4. Disconnect the cable that connects the angle sensor to the control board from the angle sensor. When you reconnect
the cable, make sure that the cable maintains the same orientation.
5. Using a #2 Phillips screwdriver and a 3/8” combination wrench, remove the two screws and the two nylon nuts that
secure the angle sensor support and the retaining plate to the frame. Remove and discard the angle sensor. Save the
screws and nuts.
N
No
otte
e -- If the angle sensor is glued to the support and retaining plate, the replacement part number is 80-5001.
1900-009-002 Rev AA.6