A
B
F
Fiig
gu
urre
e 1
16
6 –– F
Fo
oo
ott cco
ovve
err rre
em
mo
ovva
all
4. Place a sawhorse under the corner of the litter frame on the side of the load cell to be replaced. For hydraulic bases,
make sure that the sawhorse does not interfere with lowering the litter. For fixed bases, use a floor jack to raise the base
onto the sawhorse.
5. Lower the litter onto the sawhorse. Make sure that the litter frame is free of any tension.
6. Using a 9/16” combination wrench and a 9/16” socket, remove the two bolts and two locknuts (B) that secure the load
cell to the frame support (Figure 17). Save the bolts and locknuts.
N
No
otte
e -- Using a torque wrench, torque the bolts to 450 in-lb during reinstallation.
POSITION #1
A
B
B
D
E
G
F
D
C
F
Fiig
gu
urre
e 1
17
7 –– L
Lo
oa
ad
d cce
ellll
7. Turn the load cell to disconnect the cable. Slide the load cell out.
8. Using diagonal pliers, cut cable ties that secure the load cell cable to the litter frame.
EN
30
1900-009-002 Rev AA.6