![Stryker 7600-000-050 Скачать руководство пользователя страница 33](http://html.mh-extra.com/html/stryker/7600-000-050/7600-000-050_maintenance-manual_1385854033.webp)
N
Nu
urrsse
e cco
on
nttrro
oll p
pe
en
nd
da
an
ntt rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Diagonal pliers
P
Prro
occe
ed
du
urre
e::
1. Apply the brakes.
2. Raise the product to the highest height position.
3. Power down the bed and unplug the power cord from the wall outlet.
4. Using diagonal pliers, cut and discard the cable tie that secures the fowler (backrest) actuator cable to the wire routing
channel.
5. Remove the three cable routing covers that secure the gatch (legrest) actuator cable to the litter frame.
6. Unplug the nurse control pendant cable from the junction box.
7. Dispose of the nurse control pendant.
8. Reverse steps to reinstall.
9. Verify proper operation of the product before returning it to service.
S
Siid
de
erra
aiill cco
on
nttrro
oll p
pa
an
ne
ell ((iin
nssiid
de
e a
an
nd
d o
ou
uttssiid
de
e ssiid
de
erra
aiill)) rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Flat screwdriver
P
Prro
occe
ed
du
urre
e::
1. Apply the brakes.
2. Raise the product to the highest height position.
3. Unplug the power cord from the wall outlet.
4. Using a flat screwdriver, carefully push in and up in the slotted access hole to remove the control panel from the siderail
hoop.
5. Unplug the cable connecting the siderail control panel to the junction box.
6. Remove and discard the siderail control panel.
7. Reverse steps to reinstall.
8. Verify proper operation of the product before returning it to service.
C
Co
on
nttrro
oll b
bo
oxx rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
Diagonal pliers
•
Small flat screwdriver
•
3/8 in. drive ratchet
•
10 mm socket
•
10 mm combination wrench
P
Prro
occe
ed
du
urre
e::
1. Apply the brakes.
2. Raise the product to the highest height position.
3. Power down the bed and unplug the power cord from the wall outlet.
SK-7600 REV 06
29
EN
Содержание 7600-000-050
Страница 2: ......
Страница 16: ...S Se er riia all n nu um mb be er r llo oc ca at tiio on n A EN 12 SK 7600 REV 06 ...
Страница 21: ...A B C D E F SK 7600 REV 06 17 EN ...
Страница 68: ...H HM M 1 17 7 3 31 14 4 A A A HM 17 314 1 EN 64 SK 7600 REV 06 ...
Страница 69: ...H HM M 1 17 7 1 16 6 A A HM 17 16 1 SK 7600 REV 06 65 EN ...
Страница 70: ...H HM M 1 17 7 1 15 56 6 A A A HM 17 156 1 EN 66 SK 7600 REV 06 ...
Страница 71: ...H HM M 1 17 7 3 31 10 0 H HM M 1 17 7 3 31 11 1 A HM 17 310 HM 17 311 1 SK 7600 REV 06 67 EN ...
Страница 72: ...H HM M 1 17 7 3 31 13 3 A A HM 17 313 1 EN 68 SK 7600 REV 06 ...
Страница 73: ...H HM M 1 17 7 3 31 12 2 A A HM 17 312 1 SK 7600 REV 06 69 EN ...
Страница 74: ...H HM M 1 17 7 3 30 04 4 A B A HM 17 304 1 B HM 17 304 1 EN 70 SK 7600 REV 06 ...
Страница 75: ...H HM M 1 17 7 3 30 03 3 A A A HM 17 303 1 SK 7600 REV 06 71 EN ...
Страница 76: ...H HM M 1 17 7 3 30 05 5 A A HM 17 305 1 EN 72 SK 7600 REV 06 ...
Страница 77: ......
Страница 78: ...Stryker Medical 3800 E Centre Avenue Portage MI 49002 USA SK 7600 REV 06 WCR AA 11 2021 08 ...